| Fluoroscopic Capacity | |
| Max rated capacity | Tube Current 4mA, Tube Voltage 110kV |
| Automatic Fluoroscopy |
Tube Voltage:40kV~110kV adjust automatically
Tube Current:0.3mA~4mA set manually |
| Manual Fluoroscopy |
Continuous Tube Voltage:40kV~110kV
ContinuousTube Current:0.3mA~4mA |
| PulseFluoroscopy: |
Continuous Tube Voltage:40kV~110kV
Continuous Tube Current:4.1mA~8mA |
| PhotographyCapacity | |
| Max rated capacity | 3.5 KW |
| Tube Voltage and Current Combination |
40kV~49kV 1 ~125 mAs
50kV~59kV 1 ~110 mAs 60kV~69kV 1 ~90 mAs 70kV~79kV 1 ~80 mAs 80kV~89kV 1 ~71 mAs 90kV~99kV 1 ~63 mAs 100kV~110kV 1 ~40 mAs |
| Plate holder Size | 200mm250mm(8''10'')or 250mm300mm(10''12'') |
| X-ray Tube | |
| X-ray Tube Special for High Frequency | Fixed anode X-ray tube with 2 foci:Large focus: 1.5mm, small focus: 0.6mmInverter Frequency: 40KHzThermal capacity: 30KJ (40HU) |
| Video System | |
| Image Intensifier | Image Intensifier made by TOSHIBA (9'', 6'', 4.5'') |
| CCD Vidicon | |
| Monitor |
Horizontal 1000 lines and vertical 800 lines,
Bandwidth: 12.5MHz, Image/sec: 25 |
| CCU(central control) | Recursive Filter: K=8, 7 images storage, image upright, image overturnpositive & negative image; LIH(last image freeze), and OSD(monitor display) |
| Structure | |
| Direction wheel | 90revolution, can change the moving direction of the unit. |
| Ascending & Descending Range of Pillar | 400mm |
| C-arm | Forward and Backward Movement: 200mmRevolution around Horizontal Axis: 180Revolution around Vertical Axis:15Slip on orbit: 120(+90~ -30) |